For the above operating procedures, the corresponding SCPI commands should be:
MODE CCH Sets the mode
TRAN ON Enables the transient operation
CURR:LOW 1 Sets a value to low level
CURR:HIGH 5 Sets a value to high level
TRAN:HTIME 200ms Sets a value to high level time
TRAN:RTIME 10us Sets a value to time for rising edge
TRAN:FTIME 10us Sets a value to time for falling edge
TRAN:MODE PULS Selects pulsed transient operation
INPUT ON Turns on the load
.
.
Trig One trigger occurs
4.7.3 Toggled Transient Operation
Example 1: Assume that the load is in external triggering mode, set the load resistance value to periodically switch between 200 Ω and
500 Ω; set the time for rising edge (TimeR) to 10 ms; set time for falling edge (TimeF) to 10 ms; the load is in toggled transient opera-
tion, and a trigger occurs at this time.
For the above operating procedures, the corresponding SCPI commands should be:
MODE CRH Sets the mode
TRAN ON Enables the transient operation
RES:LLEV 200 Sets a value to low level
RES:HLEV 500 Sets a value to high level
TRAN:RTIME 10us Sets a value to time for rising edge
TRAN:FTIME 10us Sets a value to time for falling edge
TRAN:MODE TOGG Selects toggled operation
INPUT ON Turns on the load
.
.
Trig One trigger occurs
Procedure Description Display
1
Press the key to enter into the mode selection and parameter setting menu; use the ▲ ▼ keys to select CRH
mode; press the Enter key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
20.000Ω CRH OFF
2Press the Tran key to enter into transient operation. The display shows “t” to indicate in CRH status.
0.000V 0.000A
2000.0Ω tCRH OFF
3 Press the key to enter into the transient operation menu.
►LevelL: 2000.0Ω
LevelH: 2000.0Ω
4 Use Entry keys or use the knob together with ◄ ► keys to set LevelL to 200, and press the Enter key to confirm.
LevelL: 200.00Ω
►LevelH: 2000.0Ω
5 Use the Entry keys or the knob together with ◄ ► keys to set LevelH to 500, and press Enter key to confirm.
►TimeL: 400.00ms
TimeH: 200.00ms
6
Press the ▼ key to select TimeR; Use the Entry keys or the knob together with ◄ ► keys to set TimeR to 10 ms, and
press the Enter key to confirm.
TimeR: 10.00ms
►TimeF: 10.00ms
7 Use the Entry keys or the knob together with ◄ ► keys to set TimeF to 10 ms, and press Enter key to confirm.
►Mode: ◄ Puls ►
8Use ◄ ► keys or knob to set Mode value to Togg, and press the Enter key to confirm. ►Mode: ◄ Togg ►
9 Press the key to exit the transient operation menu.
0.000V 0.000A
200.00Ω tCRH OFF
10 Press the key to turn on the load.
0.000V 0.000A
200.00Ω tCRH
11 One trigger occurs when pressing + key or the signal at the trigger input terminal (TRIG IN) is low level.
0.000V 0.000A
500.00Ω tCRH ON