For the above operating procedures, the corresponding SCPI commands should be:
MODE CV Sets the mode
TRAN ON Enables the transient operation
VOLT:LOW 1 Sets a value to low level
VOLT:HIGH 5 Sets a value to high level
TRAN:LTIME 400 ms Sets a value to low level time
TRAN:HTIME 200 ms Sets a value to high level time
TRAN:RTIME 10 ms Sets a value to time for rising edge
TRAN:FTIMR 20 ms Sets a value to time for falling edge
TRAN:MODE CONT Selects continuous transient operation
INPUT ON Turns on the load
4.7.2 Pulsed Transient Operation
Example 1: Assume that the load is in external triggering mode, set the load current value to periodically switch between 1 A and 5 A;
set the time for rising edge (TimeR) to10 ms; set high level time (TimeH) to 200 ms; set time for falling edge (TimeF) to 10 ms; the load
is in pulsed transient operation, and a trigger occurs at this time.
Procedure Description Display
1
Press the key to enter into the mode selection and parameter setting menu; use the ▲ ▼ keys to select CCH
mode; press the Enter key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
0.000A CCH OFF
2Press the Tran key to enter into transient operation. The display shows “t” to indicate in CCH status.
0.000V 0.000A
0.500A tCCH
3 Press the key to enter into the transient operation menu.
LevelL: 0.500A
LevelH: 1.000A
4 Use Entry keys or use the knob together with
keys to set LevelL to 1, and press the Enter key to confirm.
LevelL: 1.000A
LevelH: 1.000A
5 Use Entry keys or use the knob together with
keys to set LevelH to 5, and press the Enter key to confirm.
TimeL: 400.00ms
TimeH: 400.00ms
6
Press the key to select TimeH; use Entry keys or use the knob together with
keys to set TimeH to 200 ms, and
press the Enter key to confirm.
TimeR: 10.00ms
TimeF: 20.00ms
7 Use Entry keys or use the knob together with
keys to set TimeR to 10 ms, and press the Enter key to confirm.
TimeR: 10.00ms
TimeF: 20.00ms
8 Use Entry keys or use the knob together with
keys to set TimeF to 10 ms, and press the Enter key to confirm. Mode: Cont
9Use the keys or knob to set Mode value to pulse, and press the Enter key to confirm. Mode: Puls
10 Press the key to exit the transient operation menu.
0.000V 0.000A
1.000A
tCCH OFF
11 Press the key to turn on the load.
0.000V 0.000A
1.000A
tCCH ON
12 One trigger occurs when pressing + key or the signal at the trigger input terminal (TRIG IN) is low level.
0.000V 0.000A
5.000A
tCCH ON