User manual
GMC-I Messtechnik GmbH 37
4.7 Transient Operation
The operating procedures for transient operation are shown below:
1 Press the
key to enter into the mode selection and parameter setting menu; use ▲ and ▼ keys to select one basic operating
mode; press the Enter key to confirm and exit the mode selection and parameter setting menu.
2 Press the
key to enter into transient operation. The display shows “t” to indicate in the basic test mode.
3 Press the key to enter into the transient operation menu.
4 Use ▲
and ▼ keys to select the parameter: LevelL, LevelH, TimeL, TimeH, TimeR und TimeF
5 Use Entry keys or use the knob together with ◄ ► keys to set values for LevelL, LevelH, TimeL, TimeH, TimeR, and TimeF.
Use
◄ ► keys or knob to set Mode value: continuous, pulsed or toggled operation.
6 Press the key to exit the transient operation menu.
7 Press the key to turn on the load.
8 If it is the pulsed transient operation or the toggled transient operation, one trigger occurs when pressing key +
key or
the signal at the trigger input terminal (TRIG IN) is low level.
4.7.1 Continuous Transient Operation
In continuous transient operation, the load switches at defined intervals between high and low level.
Relevant parameters:
– low-Pegel (LevelL),
– high-Pegel (LevelH),
– low-Pegel-Zeit (TimeL),
– high-Pegel-Zeit (TimeH),
– time for rising edge (TimeR) and
– time for falling edge (TimeF).
Example 1: Set the load voltage value to periodically switch between 1 V and 5 V; set the time for rising edge (TimeR) to 10 ms; set
high level time (TimeH) to 200 ms; set time for falling edge (TimeF) to 20 ms; set low level time (TimeL) to 400 ms, and the load is in
continuous transient operation. The operating procedures are shown as below:
Procedure Description Display
1
Press the key to enter into the mode selection and parameter setting menu; use ▲ ▼ keys to select CV
mode;
press the Enter key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
0.000
V CV OFF
2Press the Tran key to enter into transient operation. The display shows “t” to indicate in CV mode.
0.000V 0.000A
80.000V
tCV OFF
3 Press the key to enter into the transient operation menu.
► LevelL: 80.000V
LevelH: 80.000V
4 Use Entry keys or use the knob together with
◄ ► keys to set LevelL to 1, and press the Enter key to confirm.
LevelL: 1.000V
►LevelH: 80.000V
5 Use Entry keys or use the knob together with
◄ ► keys to set LevelH to 5, and press the Enter key to confirm.
►TimeL : 530.00ms
TimeH : 500.00ms
6 Use Entry keys or use the knob together with
◄ ► keys to set TimeL to 400 ms, and press the Enter key to confirm.
TimeL : 400.00ms
►TimeH : 500.00ms
7
Use Entry keys or use the knob together with
◄ ► keys to set TimeH auf 200 ms, and press the Enter key to confirm.
The menu will show the next item automatically.
►TimeR: 100.00ms
TimeF: 100.00ms
8 Use Entry keys or use the knob together with
◄ ► keys to set TimeR auf 10 ms, and press the Enter key to confirm.
TimeR: 10.00ms
►TimeF: 100.00ms
9 Use Entry keys or use the knob together with
◄ ► keys to set TimeF auf 20 ms, and press the Enter key to confirm. ►Mode: ◄ Cont ►
10
Betätigen Sie die Taste
◄ ► bzw. den Drehgeber, um Cont zu setzen
und bestätigen Sie mit Enter.
►Mode: ◄ Cont ►
11 Use keys or knob to set Mode value to Cont, and press key to confirm.
0.000V 0.000A
1.000V
tCV OFF
12 Press the key to turn on the load.
0.000V 0.000A
1.000V
tCV ON
0.000V 0.000A
5.000V
tCV ON