4.6 Short Circuit Operation
The operating procedures for short circuit operation are shown below:
Example 1: The load was in CV mode previously and it is turned off. Set the load to short circuit in CCH status. Turn on the load to exe-
cute short circuit operation.
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT ON Sets the load to short circuit operation
MODE CCH Sets the mode
INPUT ON Turns on the load
Example 2: Based on the last example, exit the short circuit operation, and turn off the load.
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT OFF Sets the load to exit the short circuit
INPUT ON Turns on the load
Procedure Description Display
1 Press the key to enter into main menu, and select Short (short circuit) item with the
▼ key.
Short :
On *Off
2Use the
◄ key or knob to set On, and press the Enter key to confirm.
Short :
On *Off
3 Press the key to exit the main menu. The display shows “s” to indicate in the basic test mode.
0.000V 0.000A
0.000V sCV OFF
4
Press the key to enter into the mode selection and parameter setting menu; use
▲ and ▼ keys to select one basic
operating mode; press the Enter key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
44.000A sCCH OFF
5 Press the key to turn on the load.
0.000V 0.000A
44.000A sCCH ON
Procedure Description Display
1 Press the key to enter into main menu, and select Short (short circuit) item with the
▲ key.
Short :
On *Off
2Use the
◄ key or knob to set Off, and press the Enter key to confirm.
Short :
On *Off
3 Press the key to exit the main menu. On the display, “s” disappears in CCH status.
0.000V 0.000A
0.000
A CCH ON
4 Press the key to turn off the load.
0.000V 0.000A
0.000
A CCH OFF