SSL ELECTRONIC LOAD SERIES 32 EL
GMC-I Messtechnik GmbH
15
2.3.5 Setting Address (0~254)
This function is used to set the device address of the load. To communicate with the computer or other equipment,
the SSL KONSTANTER 32 EL load must be assigned an address.
When you select Address set function, the LCD will display:
Set Address=***
New=
Set the address by using the number keyboard or the
rotary knob, and then press “OK” for confirmation. The
valid address range is from 0 to 254.
2.3.6 Enable/disable the Rotary knob
This function is to enable or disable the rotary knob.
Enable Def. *
Disable
Select the desired setting by using the UP and DOWN keys
or the rotary knob, then press “OK” for confirmation.
2.3.7 Program
SSL KONSTANTER 32 EL Series electronic load provides a program mode for dynamic testing. The user can program
the SSL KONSTANTER 32 EL Series electronic load and define the load values (current, power, resistance) and duration
of each program step. To program the SSL KONSTANTER 32 EL, please follow these 4 steps:
1、Select one working mode: constant current, constant power, constant resistance;
2、Set the length of the program (number of steps);
3、Set the load value (current, power, or resistance value) and duration time of each step;
4、Select program running mode: one time or repeat.
For example, you need to set the following program:
Constant Current mode:
Step1 1A for 2 seconds
Step2 2A for 5 seconds
Step3 3A for 10 seconds
Repeat running step1 to step3.