User manual
GMC-I Messtechnik GmbH 85
8.2 Queriable Functions and Parameters
Read out via interface Response string for remote operation
(example):
Resp.
string
length
Expla-
nation
in
chap.
on
page
Front panel
1)
RS 232C
IEEE 488
10 characters 10 characters 10 characters
Measured Value Query
I Max. measured current value from Min-Max memory [A] X
p. 28
IMAX? XXIMAX +045. 440 13
p. 66
I_ Min. measured current value from Min-Max memory [A] X
p. 28
IMIN? XXIMIN -000. 010 13
p. 66
Iout LED Momentary measured current value [A] X
p. 24
IOUT? + display XXIOUT +031. 510 13
p. 67
CV/CC/Pmax LED Momentary output operating mode (control mode)
X
2)
p. 16
MODE? CV = constant voltage, CC = constant current X X MODE CV 8
p. 68
OL = overload, OFF = output deactivated
Pout + Display LED Momentary measured power value [W] X
p. 24
POUT? XXPOUT +0662 .7 12
p. 70
U
Max. measured voltage value from Min-Max memory [V] X
p. 28
UMAX? XXUMAX +021. 310 13
p. 75
U_ Min. measured voltage value from Min-Max memory [V] X
p. 28
UMIN? XXUMIN +021. 280 13
p. 75
Uout + Display LED Momentary measured voltage value [V] X
p. 24
UOUT? XXUOUT +021. 300 13
p. 76
Function-Specific Device Queries
DELAY + Display LED Selected delay time for OCP function X
p. 24
DELAY? XXDELAY 12.0 0 11
p. 65
DISPLAY? On-off status of the 7-segment displays X X DISPLAY O N 11
p. 65
FSEt Sequence function parameter X
p. 31
FSET? XXFSET NF 9
p. 65
Ilim + Display LED Selected current setting range limit [A] X
p. 23
ILIM? XXILIM +050. 000 13
p. 66
Iset + Display LED Selected current setpoint value [A] X
p. 22
ISET? XXISET +048. 000 13
p. 67
UI_
Min-max memory enable status X
p. 27
MINMAX? XXMINMAX ON 10
p. 67
OCP Selected current limiting function X
p. 26
OCP? XXOCP OFF 7
p. 68
OUTPUT LED Output on-off status
X
2)
p. 23
OUTPUT? XXOUTPUT ON 10
p. 69
OVset + Display LED Selected overvoltage protection trigger value [V] X
p. 24
OVSET? XXOVSET +050 .0 12
p. 69
Pon Selected power-on status X
p. 27
POWER_ON? XXPOWER_ON R ST 12
p. 70
reP Selected number of repetitions for the SEQUENCE function X
p. 33
REPETITION? XXREPETITION 000 14
p. 71
READY LED
4)
SEQUENCE function status
X
p. 36
rcl + rrEP
p. 30
SEQUENCE? XXSEQUENCE H OLD,255,11
1
21
p. 71
Strt Selected start and stop memory addresses for the SEQUENCE function
X
3)
p. 32
StoP
p. 32
START_STOP? XXSTART_STOP 020,115 18
p. 72
STORE?
Recall data from start address to stop address from SEQUENCE memory
5)
XX
STORE 014, +015.500,+ 003.000,09
p. 40
STORE? n Recall data from address n from SEQUENCE memory
p. 45
STORE? n1,n2
Recall data from address n1 to address n2 from SEQUENCE memory
STORE? n1,n2,txt
Recall data from address n1 to address n2 from SEQUENCE memory with tab
37
6)
p. 72
n, n1, n2: 11, 12, ... , 255 (memory address), n2 n1, optional
txt: tab (optional)
.70, NC
tdEF Selected default time for the SEQUENCE function X
p. 31
TDEF? XXTDEF 10.00 10
p. 73
trG Selected trigger input function X
p. 29
T_MODE? XXT_MODE OUT 10
p. 74
tSEt Selected dwell time for the SEQUENCE function X
p. 31
TSET? XXTSET 00.10 10
p. 74
Ulim + Display LED Selected voltage setting range limit [V] X
p. 23
ULIM? XXULIM +035. 000 13
p. 75
Uset + Display LED Selected voltage setpoint value [V] X
p. 22
USET? XXUSET +021. 300 13
p. 76