User manual
GMC-I Messtechnik GmbH 83
8 Appendix
8.1 Adjustable Functions and Parameters
Setting saved with *SAV n:
Expla-
nation
in
chap.
on
page
Manual setting or via interface
n=1...10
n=11...255
Setting Command Significance / Effect
Manual
IEEE 488
RS 232C
Default setting
after RESET
*RST:
Function-Specific Commands and Device Settings
Addr n Set device address for RS 232 or IEEE 488 (interface configuration) X unchanged
p. 38
bAUd txt Set transmission speed (interface configuration) X unchanged
p. 39
txt: 50; 75; 150; ... ; 4800; 9600; 19.2t (transmission speed in bits per sec)
dbit txt Set number of data bits (interface configuration) X unchanged
p. 39
txt: 7; 8
DELAY v Output off delay for OCP
X
X 00.00
p. 24
v: 0; 0.01; ... , 99.99 (delay time in seconds)
XX
p. 65
DISPLAY txt Activate / deactivate the 7-segment displays
XX ON
p. 65
txt: ON, OFF (display on, display off = shorter measuring and setting rate via the interface)
FSET txt Sequence function parameter
X
X
NF
p. 65
txt NF / RU / RI
XX
ILIM v Current setting limit (soft-limit)
X
XI
nom
p. 23
v: 12.00 (example: 12.00 A)
XX
p. 66
ISET v Current setpoint
X
X X 000.000
p. 22
v: 10.75 (example: 10.75 A)
XX
p. 67
OCP txt Response selection for current limiting (overcurrent protection)
X
XOFF
p. 26
txt: OFF (limiting by means of current regulation)
XX
p. 68
ON (limiting by deactivating the output after DELAY time elapses)
OUTPUT txt Output On/Off
X
XOFF
p. 23
txt: OFF, ON (output deactivated, output activated)
XX
p. 69
OVSET v Overvoltage protection trigger value
X
X OVSETmax
p. 24
v: 32.5 (example: 32.5 V)
XX
p. 69
Pbit txt Set parity bit (interface configuration) X unchanged
p. 39
txt: nonE no parity bit
ZEro always 0
EVEn even numbered
odd odd numbered
onE always 1
Pon txt Device setting selection after power-up
X
unchanged
p. 27
POWER_ON txt txt: RST (default setting)
XX
p. 70
RCL (last used settings)
SBY (last used setting, output deactivated)
rEP n Set number of repetitions for the SEQUENCE function
X
X unchanged
p. 33
REPETITION nn:0 (continuous repetition)
XX
p. 71
1; 2; ... , 255 (number of sequence repetitions)
rnd txt Round off the displayed measured value (setup) X unchanged
p. 28
txt: 0; –1; –2 (no rounding, round by one or two decimal places)
Sbit txt Set number of stop bits (interface configuration) X unchanged
p. 40
txt: 1; 2
SEq txt Sequence, automatic memory recall
X
running sequence
STOP
p. 34
SEQUENCE txt txt: Go (start); (off, on (close / end, open))
XX
p. 71
hold (pause/suspend); cont (resume); strt, step, stop (step-by-step control)
Strt m Start and stop memory address for the SEQUENCE function
X
X unchanged
p. 32
StoP n m: 11; 12; ... , 255 (start address = beginning of sequence)
p. 32
START_STOP m, n n: 11; 12; ... , 255 (stop address = end of sequence, n m)
XX
p. 72
STORE n, v1, v2, v3, txt Direct transfer of parameter data to memory X X no effect
p. 72
n: 11; 12; ... ; 255 (memory address)
v1:
voltage setpoint
v2:
current setpoint
v3:
memory location-specific dwell time
txt: NC: no function, CLR: complete deletion of memory location data; NF / Ru / RI
TDEF v Set default time for the SEQUENCE function
X
X
unchanged
p. 32
(TDEF is used instead of TSET if TSET has not been set to any specific value, or if it has been set to 0.)
XX
p. 73
v: 0.01; 0.02; ... , 99.99 (delay time in seconds)