User manual
8 GMC-I Messtechnik GmbH
1.5 Technical Data
1.5.1 General Data
Output
Regulator type Primary switched-mode regulator
Operating modes Adjustable constant voltage / constant
current source with automatic sharp
transition
Output insulation Floating output with “safe electrical
separation” from the mains input and
computer interfaces
Max. allowable potential, output to ground:
120 V
Capacitance, output to ground (housing)
Series 62 N: 500 W / 1000 W: typ. 90 nF
Series 64 N: 2000 W / 3000 W: typ. 180 nF
Output operating range
Analog Interface
Functions – Auto-sensing mode
– Programmable trigger input
– Voltage control input (0 ... 5 V)
– Current control input (0 ... 5 V)
– Voltage monitor output (0 ... 10 V)
– Current monitor output (0 ... 10 V)
– Master-slave parallel operation
– Master-slave series operation
IEC 625 – IEEE 488 interface (common option with RS 232, variant 2)
Interface functions
(see chapter 6.2 for description of functions.)
SH1 – SOU RCE HANDSHAKE
AH1 – AC CEPTOR HANDSHAKE
T6 – TALKER
L4 – LISTENER
TE0 No extended talker function
LE0 No extended listener function
SR1 – SER VICE REQUEST
RL1 – REMOTE / LOCAL
DC1 – D EVICE CLEAR
PP1 – PARALLEL POLL
DT1 – D EVICE TRIGGER
C0 – No controller function
E1 / 2 – Open collector driver
Codes / formats per IEEE 488.2
Max. setting rate approx. 40 settings per second
Max. sampling rate approx. 15 measurements per second
V.24 – RS 232C Interface (optional, variant 1 or 2)
Transmission mode half-duplex, asynchronous
Baud rate 110 ... 19,200 baud, adjustable
Codes / formats Per IEEE 488.2
Max. setting rate Approx. 2 settings per second
Max. sampling rate Approx. 2 measurements per second
Power Supply
Line voltage 62 N: 230 V ~ +10 / –15%,
47 … 63 Hz
64 N: 3 x 400 / 230 V ~ +10 / -15%,
47 … 63 Hz
Inrush Current Max. 50 A
S
Fuse 62 N: 1 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
64 N: 3 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
Electrical Safety
Safety class I
Overvoltage
category II for mains input
I for output and interfaces
Fouling factor 2
Earth leakage
current 62 N: <3 mA
eff
64 N: <1 mA
eff
Electrical Isolation Rated Voltage Test Voltage
Output – mains 280 V
eff
4 kV ~ (type test)
Output – bus /
ground 120 V
S
1.5 kV ~
Mains – bus /
ground 230 V
eff
2.2 kV –
Bus – ground No electrical isolation
IEC 61010-1: 1990 + A1: 1992
DINEN61010-1: 1993
VDE 0411-1: 1994
DIN VDE 0160: 1988 + A1: 1989 class W1
VDE 0805: 1990
EN 60950: 1992
Protection IP 00 at device terminals and
interface terminals
IP 20 at housing
Electromagnetic Compatibility
Product standard EN 61326-1: 1997 + A1: 1998
Interference
emission EN 55022: 1998 class A
Interference
immunity EN 61000-4-2: 1995, feature B
EN 61000-4-3: 1996 + A1: 1998, feature A
EN 61000-4-4: 1995, feature C
EN 61000-4-5: 1995, feature B
EN 61000-4-6: 1996, feature B
EN 61000-4-11: 1994, feature A
Ambient Conditions
Climatic category KYG per DIN 40 040
Temperature range Operation: 0 to 40° C
Storage: –20 to +70° C
Relative humidity Operation: 75%,
no condensation allowed
Storage: 65%
Cooling With integrated fan
(2-step temperature control)
Inlet vent: side panel
Outlet vent: rear panel
Operating noise Noise pressure level at a distance of 30 cm
with fan set to low / high
62 N: 64 N:
front 18 / 28 dBA 20 / 30 dBA
rear 23 / 35 dBA 26 / 38 dBA
left and right 20 / 30 dBA 26 / 36 dBA
U / V
U
nominal
0,5
P
nominal
P
typ
P
dyn
(< 5 ms)
0.5 I
nominal
I
nom
0
I / A
Voltage
Setting Range
Current
Setting Range
P
typ
1.3 x P
nominal
U
nominal