User manual

8 GMC-I Messtechnik GmbH
1.5 Technical Data
1.5.1 General Data
Output
Regulator type Primary switched-mode regulator
Operating modes Adjustable constant voltage / constant
current source with automatic sharp
transition
Output insulation Floating output with “safe electrical
separation” from the mains input and
computer interfaces
Max. allowable potential, output to ground:
120 V
Capacitance, output to ground (housing)
Series 62 N: 500 W / 1000 W: typ. 90 nF
Series 64 N: 2000 W / 3000 W: typ. 180 nF
Output operating range
Analog Interface
Functions – Auto-sensing mode
– Programmable trigger input
– Voltage control input (0 ... 5 V)
– Current control input (0 ... 5 V)
– Voltage monitor output (0 ... 10 V)
– Current monitor output (0 ... 10 V)
– Master-slave parallel operation
– Master-slave series operation
IEC 625 – IEEE 488 interface (common option with RS 232, variant 2)
Interface functions
(see chapter 6.2 for description of functions.)
SH1 SOU RCE HANDSHAKE
AH1 AC CEPTOR HANDSHAKE
T6 TALKER
L4 LISTENER
TE0 No extended talker function
LE0 No extended listener function
SR1 SER VICE REQUEST
RL1 REMOTE / LOCAL
DC1 D EVICE CLEAR
PP1 PARALLEL POLL
DT1 D EVICE TRIGGER
C0 No controller function
E1 / 2 Open collector driver
Codes / formats per IEEE 488.2
Max. setting rate approx. 40 settings per second
Max. sampling rate approx. 15 measurements per second
V.24 – RS 232C Interface (optional, variant 1 or 2)
Transmission mode half-duplex, asynchronous
Baud rate 110 ... 19,200 baud, adjustable
Codes / formats Per IEEE 488.2
Max. setting rate Approx. 2 settings per second
Max. sampling rate Approx. 2 measurements per second
Power Supply
Line voltage 62 N: 230 V ~ +10 / –15%,
47 … 63 Hz
64 N: 3 x 400 / 230 V ~ +10 / -15%,
47 … 63 Hz
Inrush Current Max. 50 A
S
Fuse 62 N: 1 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
64 N: 3 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
Electrical Safety
Safety class I
Overvoltage
category II for mains input
I for output and interfaces
Fouling factor 2
Earth leakage
current 62 N: <3 mA
eff
64 N: <1 mA
eff
Electrical Isolation Rated Voltage Test Voltage
Output – mains 280 V
eff
4 kV ~ (type test)
Output – bus /
ground 120 V
S
1.5 kV ~
Mains – bus /
ground 230 V
eff
2.2 kV –
Bus – ground No electrical isolation
IEC 61010-1: 1990 + A1: 1992
DINEN61010-1: 1993
VDE 0411-1: 1994
DIN VDE 0160: 1988 + A1: 1989 class W1
VDE 0805: 1990
EN 60950: 1992
Protection IP 00 at device terminals and
interface terminals
IP 20 at housing
Electromagnetic Compatibility
Product standard EN 61326-1: 1997 + A1: 1998
Interference
emission EN 55022: 1998 class A
Interference
immunity EN 61000-4-2: 1995, feature B
EN 61000-4-3: 1996 + A1: 1998, feature A
EN 61000-4-4: 1995, feature C
EN 61000-4-5: 1995, feature B
EN 61000-4-6: 1996, feature B
EN 61000-4-11: 1994, feature A
Ambient Conditions
Climatic category KYG per DIN 40 040
Temperature range Operation: 0 to 40° C
Storage: –20 to +70° C
Relative humidity Operation: 75%,
no condensation allowed
Storage: 65%
Cooling With integrated fan
(2-step temperature control)
Inlet vent: side panel
Outlet vent: rear panel
Operating noise Noise pressure level at a distance of 30 cm
with fan set to low / high
62 N: 64 N:
front 18 / 28 dBA 20 / 30 dBA
rear 23 / 35 dBA 26 / 38 dBA
left and right 20 / 30 dBA 26 / 36 dBA
U / V
U
nominal
0,5
P
nominal
P
typ
P
dyn
(< 5 ms)
0.5 I
nominal
I
nom
0
I / A
Voltage
Setting Range
Current
Setting Range
P
typ
1.3 x P
nominal
U
nominal