User manual

GMC-I Messtechnik GmbH 5
1 Technical Description
1.1 Features and Range of Applications
SSP KONSTANTERs (single-output system power supplies) are
manual and remote controllable DC power supplies for laboratory
and system use. Despite high output power, the power supplies
are small and lightweight.
The floating output features “safety electrical separation” from the
mains input as well as optional computer interfaces, and is
classified as a safety extra-low voltage circuit (SELV) in
accordance with VDE/IEC. Wide ranging nominal power values
are available from the voltage and current controlled output.
The devices are generally equipped with a control panel and
display, as well as an analog interface.
An additional interface can be plugged into the device for
integration into computer controlled systems. Two interface
variants are available to this end, as described in chapter 1.3.
Manual adjustment of voltage and current is accomplished by
means of two rotary knobs with selectable adjusting sensitivity.
Numerous additional functions can be activated with various keys
(see table below).
Two 4-place digital LED displays read out measured values and
settings. LEDs indicate the current operating mode, selected
display parameters and the status of device and interface
functions.
The analog interface can be used to set output voltage and
current with the help of external control voltages, and also serves
to link several devices for master-slave operation. Deactivation of
the output, control panel disabling and individual or sequential
retrieval of saved settings can be controlled via the floating
optocoupler input.
1.2 Functions
Adjustable functions:
Voltage and current setpoint values
Voltage and current limit values (soft-limits)
Activate / deactivate the output
Overvoltage protection trigger value
Overcurrent response (limiting with or without shutdown)
Delay time for overcurrent protection
Starting performance (power-on status)
Reset device settings
Save device settings
Recall device settings individually or sequentially
Control panel disabling
Function selection for trigger input
Round off measured value for display
Service request masks (SRQ masks)
1
Activate/deactivate digital displays
1
Self-triggering when mains power is switched on
1
Retrievable information:
Momentary measured voltage and current values
Minimum and maximum measured voltage and current values
Momentary output power
Current device settings (individual or complete)
Current device status (control mode, overtemperature, busy)
1
Occurred events (mains or phase failure, overtemperature,
overvoltage, overload, programming error)
1
–Device ID
1
Additional Functions
Sensor terminals protected against polarity reversal and
automatic switching to auto-sensing
Overtemperature protection
Output protected against reverse polarity
Backup battery for device settings memory
Recognition of mains or phase failure
Inrush current limiting
1)
Via computer interfaces only
1.3 Options and Accessories
Devices can be equipped with a plug-in interface module (2
variants) for integration of SSP KONSTANTERs into computer
controlled systems.
Variant 1 includes one digital computer interface:
One V.24 / RS 232C bit-serial interface
This interface is provided in minimized format only
(conductors: RxD, TxD, GND).
Variant 2 includes two digital computer interfaces:
One IEC 625 bit-parallel, byte-serial interface
This interface fulfills all requirements of the IEEE 488 standard
and is commonly designated as a GPIB (general purpose
interface bus) or an HP-IB (Hewlett Packard interface bus).
One V.24 / RS 232C bit-serial interface
This interface is provided in minimized format only
(conductors: RxD, TxD, GND).
Programming of all device functions, as well as querying of
measured and configured parameters is possible via both
interfaces. The IEC bus interface provides several additional,
specific interface functions.
The interface option is supplied as a separate plug-in module and
can be easily plugged into the device at the rear.
Included accessories:
1 set operating instructions
1 mounting kit for rack installation
1 mains power cable with earthing contact plug
(for 62 N only)
Additionally available accessories:
RS 232 bus cable
IEEE – IEEE bus cable
Device driver for LabView software
Device driver for LabWindows / CVI software
Device driver for HPVEE / VXI PnP software