StoP – Sequence Stop Address
Functions
The end of a range within which memory locations or content
will be deleted or added
•The
SEQUENCE
is always
defined by means of a start address and a stop
address.
Memory location at which a SEQUENCE will be ended
The stop address can correspond to any memory location from 11
to 255.
•The stop address is selected immediately prior to initially starting
a new SEQUENCE.
•The last selected stop address is always valid for any given
SEQUENCE, regardless of the point in time at which the
SEQUENCE is started.
The stop address for a SEQUENCE change can be selected
independent of the start address.
! The contents of the memory location with the Stop address
are executed as the last function (reP 
Setting Parameters
11 Lowest possible Stop address
255 Highest possible Stop address
Settings (see sequence settings and execution on page 35)
Procedure for selecting StoP: see also chapter 4.7 - Settings.
! The StoP display appears in the window for the SEq function,
along with a related numeric parameter.
Figure 4.7.3 e Path to Stop Address Selection
Select the memory location with the <
> key or the <
> key.
After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
! The last selected and displayed value for the numeric stop
address parameter is used as a default setting.
rEP – Sequence Repetitions
Functions
Defines the number of SEQUENCE repetitions from the start
address to the stop address.
•The SEQUENCE is run from the selected start address to the
stop address. If repetitions have been programmed, the
pointer automatically jumps from the stop address back to the
start address, and the sequence is started over again.
•The rEP value is stored to SETUP memory as a device setting
with the <SAVE> key.
Setting Parameters
cont Continuous repetition until the SEQUENCE is stopped
manually (page 36 and page 36) or by means of
computer control (page 71)
1 ... 255 Number of possible repetitions
Settings (see sequence settings and execution on page 35)
Procedure for selecting rEP: see also chapter 4.7 - Settings.
! The rEP display appears in the window for the SEq function,
along with a related numeric parameter [s].
Figure 4.7.3.5 a Path to rEP Setting
Select the desired number of repetitions with the <> key or
the <> key.
After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
! The last selected and displayed value for the numeric
parameter (number of repetitions) is used as a default setting.
Possible selections for
the stop address
From the SEq function group
Select the
stop address
FUNCTION
+
From the SEq function group
Select a value for
number of repetitions
FUNCTION
+