User manual

32 GMC-I Messtechnik GmbH
tdEF – Dwell Time Independent of Memory Location
Functions
Default dwell time for all voltage-current value pairs included in
a sequence to whose memory locations no special dwell time
has been assigned (tSEt = 00.00)
tdEF is used primarily to speed up programming if the value for
a certain dwell time is used repeatedly within a given
SEQUENCE.
Setting Range
0.01 s to 99.99 s Maximum resolution: 0.01 s
Settings (see sequence settings and execution on page 35)
Do not enter tdEF until all Uset, Iset and tset values have been
assigned to all necessary memory locations!
Procedure for selecting tdEF: see also chapter 4.7 - Settings.
! The tdEF display appears in the window for the SEq function,
along with a related numeric parameter [s].
Figure 4.7.3 c Path to tdEF Settings
Set resolution, i.e. the decimal place (cursor position) of the
numeric parameter by repeatedly pressing the <RESOL> key.
Enter dwell time based on the previously entered resolution
value using the <> key or the <> key.
After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
! The last selected and displayed value for the numeric dwell
time parameter is used as a default setting.
Strt – Sequence Start Address
Functions
The beginning of a range within which memory locations or
content will be deleted or added
•The SEQUENCE is always defined by means of a start address and a
stop address.
The address of the memory location from which the SEQUENCE
will start
The start address can correspond to any memory location from 11
to 255.
•The start address is selected immediately prior to initially starting
a new SEQUENCE.
•The last selected start address is always valid for any give
SEQUENCE, regardless of the point in time at which the
SEQUENCE is started.
The start address for a SEQUENCE change can be selected
independent of the stop address.
! The contents of the memory location with the Strt address are
executed as the first function.
Setting Parameters
11 Lowest possible Strt address
255 Highest possible Strt address
Settings (see sequence settings and execution on page 35)
Procedure for selecting Strt: see also chapter 4.7 - Settings.
! The Strt display appears in the window for the SEq function,
along with a related numeric parameter.
Figure 4.7.3 d Path to Start Address Selection
Select the desired memory location with the <> or the <>
key.
After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
! The last selected and displayed value for the Strt address is
used as a default setting.
Select the text parameter
from an options loop
ENTER
RESOL
Cursor position
Resolution of
numeric parameter
*
From the SEq function group
CE/LOCAL
FUNCTION SAVE
RCL
Uset/Iset FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
Possible selections for
the start address
From the SEq function group
Select the
start address
FUNCTION
+