User manual
GMC-I Messtechnik GmbH 31
FSEt - memory location-specific voltage or current function
Function
• This parameter determines which function is to be performed
during the transition to this memory location..
• This command has been available as from firmware version
03.004.
Setting parameters
nF SEQUENCE values USET, ISET, TSET without addi-
tional function
rU Voltage ramp from the previous to the new voltage
setpoint value, duration TSET and/or TDEF
rI Current ramp from the previous to the new current
setpoint value, duration TSET and/or TDEF
Settings
☞ Procedure for selecting FSEt: see chapter 4.7 - Settings.
! The display of the SEq function shows the value for FSEt and
the associated alphanumerical parameter.
☞ Set the parameter with key <> or <>
☞ The parameter is accepted upon acknowledging with the
<ENTER> key.
☞ After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
tSEt – Memory Location-Specific Dwell Time
Functions
Special dwell time (see also Figure 4.7.3):
• For the voltage-current value pair at a specific memory
location within a SEQUENCE
Setting Range
00.00 s The tdEF value is used.
00.01 s to 99.99 s Maximum resolution: 0.01 s
99.99 s Assignment of the specified time to
several consecutive SEQUENCE memory
locations with identical voltage and
current values
☞ The values include rise time and decay time (see also Figure
4.7.3). Edge times are only assured within a limited load
impedance range (see also Technical Data).
Settings (see sequence settings and execution on page 35)
☞ Procedure for selecting tSEt: see also chapter 4.7 - Settings.
! The tSEt display appears in the window for the SEq function,
along with a related numeric parameter [s].
Figure 4.7.3 b Path to tSEt Settings
☞ Set resolution, i.e. the decimal place (cursor position) of the
numeric parameter by repeatedly pressing the <RESOL> key.
☞ Enter dwell time based on the previously entered resolution
value using the <> key or the <> key.
☞ After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
☞ The last selected and displayed value for the numeric dwell
time parameter is used as a default setting.
Select the text parameter
from an options loop
ENTER
RESOL
Cursor position
Resolution of
numeric parameter
*
From the SEq function group
CE/LOCAL
FUNCTION SAVE
RCL
Uset/Iset FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change