7 APPENDIX Error Messages
7.1 Error Messages
When the front-panel ERROR annunciator turns on, and the power supply generates a short beep, one or more command
syntax or hardware errors have been detected. A record of up to 20 errors is stored in the power supply’s memory.
Errors are stored and retrieved in first-in-first-out (FIFO) order. The first error returned is the first error that was stored. When
you have read or cleared all errors from the queue, the ERROR annunciator turns off automatically.
If more than 20 errors have occurred, the last error stored in the queue (the most recent error) is replaced with -350, “Too
many errors”. No additional errors are stored until you read or remove errors from the queue. If no errors have occurred when
you read the error queue, the supply responds with + 0, “No error”.
The error queue is cleared after a *CLS (clear status) command has been executed. The *RST (reset) command does not
clear the error queue. All error records will be lost when the power is turned off.
Front-Panel operation:
If the ERROR annunciator is on, press the “Error” key repeatedly to read the errors stored in the queue in sequence. The
error queue is cleared when you read all errors and then the “ERROR” annunciator turns off.
Remote interface operation:
SYSTEM:ERROR? Read one error from the error queue
7.2 Execution Errors
Code Explanation
-101 Invalid character
An invalid character was found in the command string. You may have inserted a character such as #, $, or % in the command
keyword or within a parameter.
Example: OUTP:TRAC #ON
-102 Syntax error
Invalid syntax was found in the command string. You may have inserted a blank space before or after a colon in the
command header, or before a comma.
Example: VOLT:LEV ,1
-103 Invalid separator
An invalid separator was found in the command string. You may have replaced a colon, semicolon, or blank space with a
comma or you may have replaced a comma with a blank space.
Example: TRIG:SOUR,BUS or APPL P6V 1.0 1.0
-104 Data type error
The wrong parameter type was found in the command string. You may have specified a number where a string was
expected, or vice versa.
-108 Parameter not allowed
More parameters were received than expected for the command. You may have entered an extra parameter, or you added a
parameter to a command that does not accept a parameter.
Example: APPL? 10