User manual

LABKON Series
_________________________________________________________________________________________________
GMC-I Messtechnik GmbH 27
3.13.3 6 A Calibration
Press Enter key to verify and then you will see the following:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press the Clear key to
remove the wrong input value of present digit. Then press the Enter key to confirm.
3.13.4 9 A Calibration
Press Enter key to verify and then the display will show as follows:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press the Clear key to
remove the wrong input value of present digit. Then press the Enter key to confirm.
3.14 OV Calibration
OV calibration is a kind of automatic calibration without manual operation, which starts after the voltage calibration is
conducted and is shown in the following:
When the calibration is completed, the display will go back to the calibration menu.