User manual
36 GOSSEN-METRAWATT
Interface
Type RS-232, serial, acc. to DIN 19241
Baud rate 9600 bauds
Parity None
Data bits 8
Stop bit 1
Assignment of the interfaces
The 9-pin D-SUB connector for connection of the PSI module to the SECUTEST 0701S tester has the
following pin assignment:
The 9-pin D-SUB connection socket for connection to PC, bar code reader, etc., has the following
assignment:
Reference conditions
Battery voltage 6 V
± 0.5 V
Auxiliary power 9 V
± 0.5V DC or 8V ± 0.5V rectified
Ambient temperature +23
°C ± 2K
Relative humidity 50 %
± 5%
Storage temperature
Stockage temperature
− 20 °C ... +60 °C, except for batteries, paper and ribbon
Operating temperature 0
°C ... +40 °C
Humidity max. 75 % RH; no dewing
1: Enable for remote control „Plus“
2: RXD
3: TXD
4: NC
5: GROUND
6: Enable for remote
control "GROUND"
7: NC
8: NC
9: +9 V
9876
54
3
21
1: NC
2: TXD
3: RXD
4: Switching input
5: GROUND
6: +5 V
7: CTS
8: RTS
9: NC
6
789
12
3
45