User manual

GOSSEN-METRAWATT 31
How to output statistic data
The statistic data stored in the PSI module for the 8 classes can be requested by means of the fol-
lowing instruction:
STAx?
If the statistic data of defined classes is to be requested, the instruction must be completed by the
desired class. The options are given in angle brackets:
STAx?<A><;><B><;><C><;><D><;><E><;><F><;><G><;><H>
Response: Number of DUT;percentage thereof(100);number of visual errors;percentage
thereof;number R
SL
; percentage thereof;number sum ISO; percentage thereof; number
R
Iso
; number I
EA
; I
Sonde
; number ýI; number error sum;percentage thereof;class
name;identification byte$CS<CR>
The identification byte is composed of:
01: first error; not yet stored
02: all errors; not yet stored
1X: ..... errors; some stored
If several statistics are requested, in place of <CR> a ';' is sent as delimiter after each statistic.
The last statistic in a response ends with $CS<CR>
Within a statistic (of a class), the data is separated from each other by a delimiter ';'. If there is no
data available, it is omitted in the transmission, but the delimiters remain as place holders.
Example:
STA0?A;C;D
Response: 103; 100; 0; 0; 2; 1.9; 0; 0; 0; 0; 0; 2; 1.9; Class A; 11; $CHECKSUM;
; ; ; ; ; ; ; ; ; ; ; ; ; ; Class C; 01$CHECKSUM;
1; 100; 0; 0; 1; 100; 0; 0; 0; 0; 0; 0; 1; 100; Class D; 12$CHECKSUM<CR>
Line breaks and spaces in the response transmissions of the examples are for clear
presentation only, they are not included in the transmission.