User manual

30 GOSSEN-METRAWATT
How to output the contents of the protocol memory
WERx? This instruction is used to output the memory contents of the PSI module. The PSI module
sequentially sends all records of the memory. This record contains a protocol which is
generated by the SECUTEST 0701S tester, completed by the protocol entries in the PSI
module. Should the memory be empty, a Not-Acknowledge is sent.
WERTEx= SECUTEST protocol; date;time;device;manufacturer;type; identnumber;name;street,zip-
code;twon;repair1; repair2; repair3;repair4;repair5,repair6;repair7;repair8, repair9,
repair10;protocol number$CS<CR>
– If several records are stored, after the first record a ';' is sent as delimiter in place of <CR>
– The last record in a response ends with $CS<CR>
– Within a record, the data is separated from each other by the delimiter ';'. If there is no data avail-
able, it is omitted in the transmission, but the delimiters are maintained as place holders.
Note
The SECUTEST 0701S tester transmits only data that has been acquired. That is why
the number of transmitted strings may vary.
EXAMPLE: WERΓΈ?
Prompt: VALUESx=PHOENIX;2701000000000000;
0.23;1.0;5.23;0.5;534;500;0.003;7.0;1.37;3.5;10.06.1995;13:45;
Waschmaschine;Siemens;A0815;A1234567;Huber Franz;Untere Kanalstr. 28;
98234;Moosbach;Motor;Pumpe;Kondensator;;;;;;;;0001$CHECKSUMME;
PHOENIX;0602000000000000;7.34;2.00;523;500;0.23;1.0;0002$CHECKSUMME;
TESTER;0602000000000000;7.34;2.00;523;500;0.23;1.0;0003$CHECKSUMME<CR>
Line breaks and spaces in the response transmissions of the examples are for clear pres-
entation only, they are not included in the transmission.
With the first device under test, a Class I test was performed, followed by a function test,
and a protocol completely filled in;
with the 2nd device under test, only a Class II test was performed without any entries via
the keyboard.
The 3rd device under test was measured with another SECUTEST 0701S tester (name
Tester), otherwise protocol same as 2nd device under test.