With the following telegram, the SECUTEST PSI acknowledges whether it has executed an instruction
(Y=Yes) or not (N=No). An acknowledgement, for example, is also made when a reply message is not
required:
Acknowledge: Instruction executed, transmission format:
Y<ADRESSE><DELIMITER$><CHKSUM> <CR>
Not-Acknowledge: Instruction not executed, transmission format:
N<ADRESSE><DELIMITER$><CHKSUM> <CR>
9.2 Instruction set, validity and syntax
The syntax checker in the tester checks only the first three characters of each instruction. If they are
correct, the instruction is accepted.
Parameters and instructions are configured in such a way that there will be no ambiguities. Writing in
uppercase or lowercase letters is of no importance. At any rate, the checksum can only be obtained
from the actually sent characters.
The parameter x is the address of the device (SECUTEST 0701S or SECUTEST PSI) which is to be
addressed. If only one equipment is connected, an address need not be given.
Address range: 0
≤ x ≤ 90
9.3 Instruction list for the PSI module
IDN? Gives the equipment identification
ESR? Gives the equipment status
RST! Clears the entire memory including top and bottom lines, date, time and statistics
MEM! Deletes the data in the protocol memory
WER? Gives the complete memory contents that is, the data from the PRO? instruction,
completed by the entries in the PSI module
FKT? Gives the data of the function test
DAT? Gives date and time
DAT! Sets date and time
KOP? Reads the top lines
KOP! Sets the top lines 1 ... 5
FUS? Reads the bottom lines
FUS! Sets the bottom lines 1 ... 5
FOO! Sets the bottom lines 6 and 7
STA? Reads the statistic data
STA! Deletes the statistic data
CLA? Reads class name and identification byte
CLA! Changes class name and type of error
PRO? Reads the data of safety test and function test
(see operating instructions SECUTEST 0701S)
Instructions unknown to the PSI module are automatically passed to the SECUTEST 0701S
(e.g. in the case of instruction PRO?)