7 Operation with bar code reader B3261
The bar code reader B3261 (accessory) allows for all information available in bar code form to be
quickly, easily and safely entered into the test protocols. This type of data entry permits time-saving
and low-cost acquisition of larger amounts of information, e.g. series measurements of devices pro-
vided with bar codes.
7.1 Connecting the bar code reader
Connect the reader to the RS-232 interface (4) of the PSI module.
A PSI LC display window must not be active!
The bar code reader confirms the correct connection with a four-fold sound signal.
7.2 Configuring the bar code reader
For operation on the PSI module, the bar code reader B3261 (DL80) must be configured. For that
reason, codes (instructions) must be entered. This is possible
– By entering the bar code shown below or
– By entering several bar codes in the order listed in the following table. You will find these codes
on the pages "Appendixes 1" and "Appendixes 2" of the operating instructions of the bar code
reader.
The following bar code contains all information of the bar code series given in the table.
Enter as described in section 7.3, page 21.
The following bar code includes all information about the bar code line of the table.
Bar code for configuration of the reader for operation on the PSI module
7.2.1 Application of EAN13 codes (e.g. for the SE-Q.base program)
Using EAN13 codes, the 13th digit (test numeral) is disabled when reading the following bar code.