5 How to display, print and store protocol
Note
Storage of the results of the safety and function tests as well as their entry into pro-
tocol and statistics is only possible after tests according to DIN VDE 0701; that is,
the function switch must be set the "AUTO", "SK I", "SK II", "SK III" or "T 240"
position!
With the switch set to "MENU" and "FUNCTION TEST", only the display shown on the
LCD at a time can be printed out and the result of the last function test at a time can be
stored (see section 10, page 34).
You can load the result of the last test according to DIN VDE 0701 at a time into the PSI module, store
it there under an Ident Number, enter it and print it out. The protocol of the last test can be printed
out several times.
The results of a complete working day (approximately 200 protocols) can be stored in the memory of
the PSI module. Pressing the key (15) several times makes for the same record to be
overwritten again and again.
The test protocol automatically contains the test result including the measured values and limits as
well as information on the visual inspection. Further specific information on device under test, cus-
tomer and repair can be integrated in the test protocol by entries from the keyboard or via a bar code
reader (see section 7, page 20).
The protocol is presented on the LCD in several windows.
5.1 How to display protocol, enter and store texts
Request the PSI module via the key
Select Protocol and confirm with
First, the test result including measured values
and limits is shown. The display contains available data only.
In further windows that can be selected with the and
keys, you can display information on the visual inspection as
well as enter texts via the keyboard, and bar codes via the
bar code reader (see section 7, page 20). A maximum of
24 characters can be entered into one line.
End the text entry of a line by pressing the key.
At the same time, this brings you to the next line.
To store, press the key.
returns you to the PSI menu.
When storing the protocol, the successive ident number is
output between date and time.
STORE
PRINT
ENTER
STORE