Data Sheet

GMC-I Messtechnik GmbH 7
METRAHITIM XTRA & METRAHITIM E-DRIVE
Isolation Tester, Milliohmmeter,
TRMS Multimeter, Short-Circuited Coil Tester
Ambient Conditions
Accuracy range 0 C to +40 C
Operating temperatures
(Storage temperature
with batteries)
10 C ... +50 C
20 C ... +50 C with rubber holster
Storage temperatures 25 C ... +70 C (without battery)
Relative humidity 40 to 75%, no condensation allowed
Elevation To 2000 m
Deployment Indoors, except within specified ambient
conditions
Data Interface
Type
Bluetooth 4.2
Frequency band 2.402 ... 2.480 GHz
Transmitting power max. 91 mW
Functions Query measuring functions
and parameters
Query momentary measurement data
Internal Measured Value Storage
Memory capacity 64 MBit for approx. 300,000 measured
values with indication of date and time
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 235 x 105 x 56 mm (without rubber
holster)
Weight Approx. 0.7 kg with battery module
Protection Housing: IP 52 (pressure equalization by
means of the housing)
Excerpt from table on the meaning of IP
Codes
Applicable Regulations and Standards
METRAHIT IM XTRA
with Accessory COIL Adapter 50mH (Z270F)
METRAHIT IM XTRA
with Accessory COIL Adapter XTRA (Z270M)
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
0 not protected 0 not protected
1 50.0 mm dia. 1 vertically falling drops
2 12.5 mm dia. 2 vertically falling drops with
enclosure tilted 15°
3 2.5 mm dia. 3 spraying water
4 1.0 mm dia. 4 splashing water
5 dust protected 5 water jets
DIN EN 61010–1
VDE 0411-1
Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
DIN EN 61326-1
VDE 0843-20-1
Electrical equipment for measurement, control and laboratory use
– EMC requirements – Part 1: General requirements
DIN EN 60529
VDE 0470-1
Test instruments and test procedures – degrees of protection
provided by enclosures (IP code)
DIN EN 61557-1
VDE 0413-1
Electrical safety in low voltage distribution systems up to 1000 V
a.c. and 1500 V d.c. – Equipment for testing, measuring or moni-
toring of protective measures
Part 1: General requirements
DIN EN 61557-2
VDE 0413-2
Part 2: Insulation resistance
DIN EN 61557-4
VDE 0413-4
Part 4: Resistance of earth connection and equipotential bonding