Datasheet

3-349-601-03
5/9.13
GMC-I Messtechnik GmbH
CP 30 / CP 330 / CP 1100 / CP 1800
AC / DC Current Probe
CP1800CP1100CP330CP30
Applications
The CP 30, CP 330, CP 1100 and CP 1800 AC / DC current
probes have been designed for reliable and accurate, contactless
measurement of AC, DC and complex waveform currents.
Their special characteristics make these current probes high-per-
formance instruments for the measurement of inverters, swit-
ched-mode power supply units, industrial control systems and
other applications which require current measurement or wave-
form analysis.
Features
Auto Power OFF
The device is switched off automatically after approx. 10 minutes
if none of its keys have been activated during this time. Automatic
shutdown can be deactivated.
Characteristic Values
1)
CP 30: DC or AC
pk
; CP 330/CP 1100/CP 1800: DC or AC
RMS
2)
under reference conditions
Operating voltage 300 V AC
RMS
or DC
Key rdg. = reading (measured value)
CP 30 CP 330 CP 1100 CP 1800
Measuring Range
1)
30 A 30 A / 300 A 100 A / 1000 A 125 A / 1250 A
Resolution 1 mA 50 mA / 100 mA 100 mA / 500 mA 100 mA / 500 mA
Overload Capacity (60 s) 500 A 600 A 2000 A 2000 A
Temperature Coefficient 0.01 % rdg./C 0.1 % rdg./C 0.1 % rdg./C 0.1 % rdg./C
Transformation Ratio
100 mV/A
10 mV/A
1mV/A
10 mV/A
1mV/A
10 mV/A
1mV/A
Overall Accuracy 1 % rdg. 2mA 1% rdg. 50/100 mA 1% rdg. 100/500 mA 1 % rdg. 100/500 mA
Frequency Range
DC up to 20 kHz (–1 dB) DC up to 20 kHz (–3 dB) DC up to 20 kHz (–1 dB) DC up to 20 kHz (–1 dB)
Phase shift
2)
<2
Load Impendance
>100 k and
100 pF
Influence of Conductor Position
1%
2)
AC and DC current measurement
CP 30: max. 30 A
CP 330: max. 450 A
CP 1100: max. 1400 A
CP 1800: max. 1800 A
•Resolution:
CP 30: 1 mA
CP 330: 50 mA
CP 1100: 100 mA
CP 1800: 100 mA
Opening angle of current clamp brackets
:
CP 30: 25 mm
CP 330: 25 mm
CP 1100: 32 mm
CP 1800: 32 mm
High accuracy
Automatic zero balancing
Auto Power Off
Using advanced Hall effect technology