User Manual
WaveAce
6
WA-GSM-E-RevA
PLEASE NOTE THE FOLLOWING:
The design of the instrument has been verified to conform to EN 61010-1
safety standard per the following limits:
Installation (Overvoltage) Categories II (Mains Supply Connector) & I
(Measuring Terminals)
Pollution Degree 2
Protection Class I
Cooling
The instrument relies on forced air cooling with internal fans and ventilation
openings. Care must be taken to avoid restricting the airflow around the
apertures (fan holes) at the sides and rear of the oscilloscope. Ensure
adequate ventilation by leaving the required 10 cm (4 inch) minimum gap
around the sides and rear of the instrument.
CAUTION
Do not block the ventilation holes located on both sides and rear of the
oscilloscope.
CAUTION
Do not allow any foreign matter to enter the oscilloscope through the
ventilation holes, etc.
AC Power Source
100 to 240 Vrms (+/-10%) AC at 50/60/400 Hz; Installation Category: 300V
CAT II
No manual voltage selection is required because the instrument
automatically adapts to line voltage.
Maximum Power Consumption: 50 VA
Note: The instrument automatically adapts itself to the AC line input within
the following ranges:
Voltage
Range:
90 to 264 VAC
Frequency
Range:
45 to 440 Hz