User Manual

Getting Started Manual
WA-GSM-E-RevA
39
Video Triggering
Use Video Triggering to trigger on fields or lines of standard video signals.
1. Press the Type option button and select Video.
2. Press the Source option button to choose a channel input
or external input.
3. Press the Polarity option button and select (Normal)
or ( Inverted). Normal triggers on the negative edge
of the sync pulse. Inverted triggers on the positive edge of
the sync pulse.
4. Press the Sync option button and select a video sync
(Line Num, All Lines, Odd Field, Even Field). If you
select Line Num, you can turn the Adjust front panel
knob to set the appointed line number.
5. Press the Next Page option button.
6. Press the Standard option button and select the video
standard for sync and line number count (NTSC or
Pal/Secam).
7. Press the Mode option button to select Auto, Normal, or
Single mode. Use Auto mode to let the acquisition
automatically run in the absence of a trigger. Use Normal
mode when you want to see only valid triggered
waveforms (when you use this mode the oscilloscope
does not display a waveform until after the first trigger).
Use Single mode when you want the oscilloscope to
acquire a single waveform.
8. Press the Setup option button to display the Trigger
Setup menu. You can use the Trigger Setup menu to
select a coupling mode and define a Holdoff value.
Coupling modes comprise DC, AC, HF Reject, and LF
Reject. Turn the Adjust front panel knob to set the
Holdoff value.