User Manual

WaveAce
38
WA-GSM-E-RevA
Pulse Triggering
Use Pulse width triggering to trigger on aberrant pulses. You can select
how to compare the trigger pulse relative to the pulse width as follows:
Positive pulse width less than pulse width setting
Positive pulse width larger than pulse width setting
Positive pulse width equal to pulse width setting
Negative pulse width less than pulse width setting
Negative pulse width larger than pulse width setting
Negative pulse width equal to pulse width setting
1. Press the Type option button and select
Pulse.
2. Press the Source option button to choose a
channel input or external input.
3. Press the When option button to select how
to compare the trigger pulse relative to the
value selected in the Set Width option.
4. Press the Set Width option button and then
turn the Adjust front panel knob to set the
pulse width.
5. Press the Next Page option button.
6. Press the Mode option button to select Auto,
Normal, or Single mode. Use Auto mode to
let the acquisition automatically run in the
absence of a trigger. Use Normal mode when
you want to see only valid triggered
waveforms (when you use this mode the
oscilloscope does not display a waveform
until after the first trigger). Use Single mode
when you want the oscilloscope to acquire a
single waveform.
7. Press the Setup option button to display the
Trigger Setup menu.