Operating Manual
English
85
Options
a. In the operating mode, tap on the symbol
.
The Options menu appears.
Option Meaning
‘Manually/STEP Vol-
ume’
Switch to titration mode
‘Show last volumes’ Displays the last volumes titrated.
‘Help Work Mode’ Shows the help text for the mode.
Favorites
You can save settings in Favorites, in order to open them again later, e.g.
aliquot lists, volume settings.
Cleaning and disinfection
Cleaning
NOTICE
The device is not autoclavable.
The device is factory-calibrated and maintenance-free.
When the device is dirty, clean the outer surface with a moist cloth . Use
water or a mild detergent solution. Avoid acidic or aggressive cleaners.
Disassembly of the device is not permitted.
UV disinfection
The device is resistant to normal exposure to a UV disinfection lamp. The
device may change color due to the UV exposure.
UV disinfection recommendations:
Light spectrum UV-C
Wave length 220 nm ... 270 nm
Exposure time per cm
2
and intended inactivation rate 2 s ... 300 s