Datasheet
GlobTek, Inc.
www.globtek.com
186 Veterans Drive,
Northvale, NJ 07647
Tel. (201) 784-1000
Fax (201) 784-0111
Drawing Title:GTM91120-3014.5-2.5-T2, Medical/ITE Power Supply, Efficiency Level: V, Case Color/Style:
GlobTek Black, Wall Plug-in/Desktop, Regulated Switchmode AC-DC, Input Rating: 100-240V~, 50-60 Hz, IEC
60320/C8 w/ Wall Blades Option, Output Rating: 30W, 12V@2.5A, Output Configuration: 1830 mm, 18/1C +
Shield Cond, UL 1185, Female Barrel 5.5*2.1*11mm No internal Spring Clip nor Locking Notch or Equal,
Ferrite: Core Size: 14.2*28.5*8.2, GlobTek Black, GT Cord # C0750, Pin Out: Center Positive (+V), Blades/Input
Cord: R-Series Desktop Insert, Approvals: C-Tick, CB, CE, China RoHS, cRUus, Double Insulation, FCC,
GOST-R, IP40, IP52 Option, Level V, LPS, PSE, RoHS 2, SIQ, Ukraine, UL/cUL, VCCI, WEEE
Model No.GTM91120-3014.5-2.5-T2
Part No.RR9KE2500CCPCIMRVB RevD
PROPRIETARY INFORMATION
PROPRIETARY OF GLOBTEK, INC. ANY REPRODUCTION, DISCLOSURE OR USE OF THIS DRAWING, IN WHOLE OR IN PART, IS HEREBY PROHIBITED EXCEPT AS SPECIFIED IN
WRITING BY GLOBTEK, INC.
sorec / 2014.02.17 at 16:43:54 GMT
Page 2 of 11
POWER SUPPLY INFORMATION
TYPE: Wall Plug-in/Desktop
TECHNOLOGY: Regulated Switchmode AC-DC
CASE COLOR: GlobTek Black
NAMEPLATE RATED
INPUT:
100-240V~, 50-60 Hz, 1.5A
INPUT CONFIG: IEC 60320/C8 w/ Wall Blades Option
WATTS: 30.0
VOLTS OUT: 12
CURRENT OUT (Amps): 2.5
BLADE/CORD INCLUDED: R-DI:R-Series Desktop Insert
BLADE/CORD INSTALLED: R-DI:R-Series Desktop Insert
EFFICIENCY LEVEL: V
OUTPUT
CONFIGURATION
1830 mm, 18/1C + Shield Cond, UL 1185, Female Barrel 5.5*2.1*11mm No internal Spring Clip nor Locking Notch or
Equal, Ferrite: Core Size: 14.2*28.5*8.2, GlobTek Black, GT Cord # C0750
CONNECTOR PIN OUT: Center Positive (+V)
LABEL SPECS: Standard GT,L-1186
PACK SPEC: Desktop/WPI with Desktop Insert
NOTES / DEVIATIONS:
DESKTOP INSERT INSTALLED
DIMENSIONS ARE IN MM UNLESS SPECIFIED OTHERWISE.
ELECTRICAL SPECIFICATIONS
A) ELECTRICAL SPECIFICATIONS:
1. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
2. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
3. Output Regulation: +/- 5% measured at the output connector
4. Line Voltage Regulation: +/- 2% typical measured at full load
5. Output Ripple (Vp-p): +/-2% maximum, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic
capacitor connected at the end of the output connector
6. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load
7. Turn-ON Delay: 3 Seconds, maximum @ full load, nominal line
8. Hold-Up Time: 8 mS typical @ nominal input voltage and full load
9. Inrush Current: 60A maximum cold start @ 240Vac input
10. Switching Frequency: 100 KHz typical
B) PROTECTION
1. Over-Voltage: Protected with a Zener diode clamp across the output
2. Short Circuit: Hiccup with auto recovery
3. Over-load: Hiccup with auto recovery
4. Input Protection: Input line and neutral fusing
C) SAFETY
1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured
3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition)
D) OTHER:
1. MTBF: 200,000 Hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
3. Humidity: 0% to 90% relative humidity
4. Storage Temperature: -10°C to 80°C
5. Cooling: Convection