Owner's manual
INSTALLATION
INITIAL INSPECTION : As soon as the power supply
unit is unpacked inspect for any damage that may have
occured during transit. Save all packing material until in-
spection is completed. If any damage is found, notify the
carriers immediately. Our authorised representatives should
also be notified.
PHYSICAL CHECK : This check should confirm that
there are no broken knobs or connectors, that the cabinet
and panel surfaces are free of dents and scratches and
the meters are not scratched and cracked.
ELECTRICAL CHECK : The power supply unit should
be checked against electrical specifications.An in-cabinet
performance check will verify proper operation.
INSTALLATION DATA : The power supply unit is shipped
ready for bench operation. It is necessary only to connect
the unit to a rated source of power and it is ready for opera-
tion.
LOCATION : The power supply unit is naturally
cooled.Sufficient space should be kept around the unit while
in operation, so that heat sinks do not remain in confined
space or close to another heating source.The ambient tem-
perature of the area around the unit should be less than
40°C.
INPUT POWER REQUIREMENTS : The power supply
unit may be operated continuously from input voltage of 115V
/ 230V 47-63Hz power source with selectable Switch.
REPACKAGING FOR SHIPMENT : To ensure safe ship-
ment of the power supply unit, it is recommended that the
package designed for the unit be used. The original pack-
aging material is reusable. Be sure to attach a tag to the
unit specifying the owner, and the fault observed with a brief
description.
REMOVING COVER : The top cover is retained in place
by 6 self tapping screws & two handle mounting screws.
To remove cover, proceed as follows :
a) Remove the handle mounting screws.
b) Remove the self tapping screws on sides.
c) Lift the cover from rear side, slide backwards & pull.
5
PCB Components ZSDT-CT/05 PCB
=======================================================================
Ref Designator Value
========================================================================
R37 2K, MFR, 1/4W, 5%
R38 1K, MFR, 1/4W, 5%
R39 1K, MFR, 1/4W, 5%
R40 4.7K, MFR, 1/4W, 5%
R41 330K, MFR, 1/4W, 5%
R42 100E, MFR, 1/4W, 5% (I CAL)
R43* 3.9K, MFR, 1/4W, 5% (I CAL)
R44 1K, MFR, 1/4W, 5%
R45 1K, MFR, 1/4W, 5%
R46* 10K, MFR, 1/4W, 5% (V CAL)
R47* 100E, MFR, 1/4W, 5% (V CAL)
R48 2K, MFR, 1/4W, 5%
R49 Shorting Link
R50 Shorting Link
R51 10E, MFR, 1/4W, 5%
R59 10E, MOR, 2W
PRESETS
PR101 5K, PRE, LIN, (V)(DEV. DROP)
PR102 500E, PRE, LIN, (V)(V CAL)
PR103 500E, PRE, LIN, (V)(I CAL)
CAPACITORS
C1 0.1µF/100V, MP
C2 0.1µF/250VAC MKP
C3 15,000µF/50V ELE
C4 0.1µF/50V, MP 10%
C5 33µF/50V, ELE
C6 100µF/50V, ELE
C7 100µF/50V, ELE
C8 1µF/50V, ELE
C9 4.7µF/50V, ELE
C10 10µF/50V, ELE
C11 100µF/50V, ELE
C12 47µF/50V, ELE
C13 1kpF/50V, CD
C14 1kpF/50V,.CD
C15 0.1µF/50V, CD
C16 10µF/50V, ELE
C17 10µF/50V, ELE
C18 0.1µF/50V, CD
C19 220µF/50V, ELE
14