User manual
ART.: 05-900-585-00 Page 8 of 21
BRU-303-0029 Rev.nr.: E
Classification
LFM LFM Medical
Medical class: Class 1
Degree of protection
against electric shock
Class I (grounded) Equipment type: B Applied Part
Permitted current
leakage
Max 0.5 mA
Protection reliability
Resistance in earth circuit –
earth connection 0.1
Power supply 230AC 50 Hz 230AC 50 Hz
Power consumption Max. 95VA Max. 95VA
Mode of operation Continuous
Protection against risk of
explosion
Not protected Not protected
Protection against liquid
penetration
IP20 Not protected
Degree of mobility Mobile table Mobile table
Type of approval ENEC 17 MEDICAL Standard
Safety test in line with EN 60598-2-4(1997),
EN 60598
1(2004)+A11(2000)+
A12(2002)
IEC/EN 60601-1(1990) 2 Ed
+A1(1993)+A2(1995)+
A13(1996)+CORRIGENDIUM(1
994)
EMC test in line with EN
55015(2000)+A1(2001)+A
2(2002)
IEC/EN 60601-1-2(2001)
Primary fuse
No fuse
IEC 127-5X20mm, 2xT0.4AH
250V
Safety
Users of this form of lighting should be aware that the manufacturer of the product
can only accept liability for the lighting when maintenance, repairs or changes are
carried out by the manufacturer or by companies which the manufacturer has
approved to carry out such work, and when components that influence the safety of
the lighting are replaced with original spare parts. The lighting is not intended for use
in areas where there is a risk of explosion.
This product complies to IEC/EN 60601-1-2(2001), electromagnetic compability for
medical devices. Electromagnetic or other interference between this equipment and
other devices should not occur. In the event of interference, please move the
equipments apart from each other.
The power cable must only be replaced by a qualified electrical fitter or other
similarly qualified professional.
Any repairs of the arm must be performed by LUXO or LUXO approved personnel.
The arm contains high tensions springs.
Maintenance
Cleaning of product and replacement of light source is the only maintenance needed.