User Manual
Table Of Contents
- Book 1: Setup & Maintenance
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Setting Up the System
- Setup Requirements
- Given Workstation
- DataRecorder and Cradle
- Storage Space for the PillCam Capsule Box
- Main Platform Components
- Connecting the Components
- Connecting the Given Workstation
- 1. Connect the Mouse cable to the Mouse connector.
- 2. Connect the Keyboard cable to the Keyboard connector.
- 3. Connect the monitor to the Workstation.
- 4. Connect the printer to the LPT connector or to the USB connector, depending on the printer’s connection cable.
- 5. If the Workstation’s voltage setting is manual, verify that the Workstation’s voltage matches the local voltage. If it does not, call Given Customer Support.
- 6. After voltage verification, connect the power cable of the Given Workstation to the electric outlet.
- 7. Connect the power cable of the monitor to the wall electric outlet.
- Connecting the DataRecorder Cradle
- Chapter 3
- Software Installation
- Chapter 4
- Multi-User Setup
- Chapter 5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB Capsule
- PillCam SB 2 Capsule
- PillCam ESO 2 Capsule
- PillCam ESO 3 Capsule
- PillCam COLON 2 Capsule
- SensorArray DataRecorder 2
- SensorArray DataRecorder 3
- SB SensorBelt for DataRecorder 2 and DataRecorder 3
- DataRecorder 2 /2C
- Cradle DataRecorder 2
- DataRecorder 3
- Cradle DataRecorder 3
- DC Power Supply
- DataRecorder 3 Memory Card
- RAPID Software
- Guidance and Manufacturer's Declarations
- Chapter 6
- Maintenance
- Chapter 7
- Troubleshooting
Maintenance
Chapter 6 67
SensorBelt Cleaning
The SensorBelt may be machine washed after removal of the SensorBelt Insert.
Follow instructions on the care label and use a mild detergent.
The surface of the SensorBelt may be wiped with any commonly used disinfectant.
The SensortBelt Insert may be wiped gently with alcohol (up to 70%).
SensorArray Cleaning
For mild cleaning (dirt, sweat), wipe the sensors gently with alcohol (up to 70%). The alcohol
will not remove the adhesive. Since alcohol is a polar solvent, do not use lavishly, and allow to
dry for 20 minutes.
To remove adhesive from the SensorArray (not from the human body), use White Benzene.
Note
White Benzene MUST be used in a ventilated area with all precautions
defined in the manufacturer’s instructions.
Alternatively, use one of the following medical adhesive removers to remove adhesive:
• B-508 Secure Solvent
• B-202 Hollister Solvent
• B-206 Detachol Adhesive Remover
Use all precautions as defined by the manufacturer.