User Manual

Table Of Contents
Technical Description
Chapter 5 55
DataRecorder 2(C)
Guidance and manufacturer’s declaration - electronic emissions
The DataRecorder 2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the DataRecorder 2 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The DataRecorder 2 uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class B
The DataRecorder 2 is suitable for use in all
establishments including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - electronic emissions
The DataRecorder 2 is intended for use in the electromagnetic environment specified below. The customer
or the user of the DataRecorder 2 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient / burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge,
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.