User Manual
Table Of Contents
- Book 1: Setup & Maintenance
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Setting Up the System
- Setup Requirements
- Given Workstation
- DataRecorder and Cradle
- Storage Space for the PillCam Capsule Box
- Main Platform Components
- Connecting the Components
- Connecting the Given Workstation
- 1. Connect the Mouse cable to the Mouse connector.
- 2. Connect the Keyboard cable to the Keyboard connector.
- 3. Connect the monitor to the Workstation.
- 4. Connect the printer to the LPT connector or to the USB connector, depending on the printer’s connection cable.
- 5. If the Workstation’s voltage setting is manual, verify that the Workstation’s voltage matches the local voltage. If it does not, call Given Customer Support.
- 6. After voltage verification, connect the power cable of the Given Workstation to the electric outlet.
- 7. Connect the power cable of the monitor to the wall electric outlet.
- Connecting the DataRecorder Cradle
- Chapter 3
- Software Installation
- Chapter 4
- Multi-User Setup
- Chapter 5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB Capsule
- PillCam SB 2 Capsule
- PillCam ESO 2 Capsule
- PillCam ESO 3 Capsule
- PillCam COLON 2 Capsule
- SensorArray DataRecorder 2
- SensorArray DataRecorder 3
- SB SensorBelt for DataRecorder 2 and DataRecorder 3
- DataRecorder 2 /2C
- Cradle DataRecorder 2
- DataRecorder 3
- Cradle DataRecorder 3
- DC Power Supply
- DataRecorder 3 Memory Card
- RAPID Software
- Guidance and Manufacturer's Declarations
- Chapter 6
- Maintenance
- Chapter 7
- Troubleshooting
Technical Description
Chapter 5 43
PillCam COLON 2 Capsule
Properties
Physical Dimensions Length: 31.5 mm
Diameter: 11.6 mm
Weight 2.9 g
Material Biocompatible plastic
Optical # of optical heads 2
Illumination 4 white light emitting diodes on each side
Field of view 172° ISO-8600-3
Effective visibility Distance: 0–30 mm
Min. detectable
object
0.09 mm
Operational Operating time Total: 10 hours:
Chemical safety Resistant to dissolution in pH=2 to pH=8
Battery type Silver Oxide batteries
Operating
temperature
20–45
°C
Storage temperature
0–30
°C
Uplink
communication
Operating frequency 434.1 MHz
Frame rate 4–35 fps
Data rate 2.7 Mbps and 8.1 Mbps
Modulation type MSK/Digital data
Frequency band
standards
FCC CFR 47 part 15 section 15.209
ERC 70-03 Annex 1 band F1
Effective radiated
power
-48.8 dBm (measured in 120 kHz bandwidth)
Downlink
communication
Operating frequency 13.56 MHz
Receiver Bandwidth +
150 KHz
Frequency band
standards
FCC CFR 47 part 15 section 15.225
ERC 70-03 Annex 9 band F