User's Manual

Capsule Endoscopy
Given Imaging Ltd.
85
The Data Recorder LED starts blinking green.
5. Instruct the patient to take the M2A Capsule out of its
holder.
6. Ask the patient to put the M2A™ Capsule in his/her
mouth, and to hold it under the tongue WITHOUT
INGESTING.
7. If after a short adjustment period, with the M2A
Capsule in the patient’s mouth, the green light keeps on
blinking, ask the patient to ingest the capsule with a sip
of water.
Note:
The time elapsed between capsule activation and
ingestion should usually not exceed 1 minutes.
In
order to minimize the potential for
radiofrequency interference from
the device after removing it from
the blister package and until it is
ingested by the patient, the M2A™
capsule should be tested as quickly
as possible. After verification
that the M2A™ capsule is
operational – it should be
immediately ingested by the
patient.
8. In the Event Form, write down the ingestion time of the
M2A Capsule.