User's Manual
Troubleshooting
Given Imaging Ltd.
92
Data Recorder
The following section describes common problems you
might encounter while using the Data Recorder during the
Capsule Endoscopy.
Recorder Initialization
PROBLEM
During the Check-In process, after clicking Check-In, the
message
Initializing Recorder
stays on and the message
Initialization is complete
does not appear.
SOLUTION
•
Check the Connection Box connection to the mains and
verify that the Connection Box LED is on.
•
Check the Recorder data directory:
•
Verify that the Recorder data directory is E:\rdata.
•
Verify that the \rdata directory is a shared directory.
Use Windows Explorer or My computer to do that.
If it is not a shared directory, select the E:\rdata
directory, right-click it and use the sharing sub-menu.
•
Check the Connection Box-Workstation connection:
•
Verify that the LAN cable is properly connected.
•
Check the connection of the 9-pin female-female cable
to the Workstation and to the Connection Box.
•
Verify computer name: