User's Manual

Capsule Endoscopy
Given Imaging Ltd.
89
just before use. If a patient returns a Battery Pack at the end
of an examination and on the following day another
examination will take place, immediately recharge the
Battery Pack. Recharging a Battery Pack is an overnight
process.
Note:
The rechargeable Battery Pack lasts at least 2 years. After two
years of daily use or after 500 rechargings, it is recommended to
replace it.
Warning:
For charging the Battery Pack use only the Charger supplied
with the GivenĀ® Diagnostic Imaging System
!
To recharge the Battery Pack
1.
Connect the Charger male 9-pin connector to the Battery
Pack 9-pin Female connector.
2.
Connect the Charger to the electric socket.
The Charger indicator is on.
When the Charger indicator blinks, the Battery Pack is
fully charged.
3.
Unplug the Charger from the socket and from the Battery
Pack.