The template has 9 holes. Eight holes indicate the
recommended location of each one of the eight sensors and
the additional hole indicates the navel as a reference point.
Place the template on the patient’s abdomen and use the
marker to make a small dot at each location.
Remove the template and locate each sensor over the dot.
Securely attach the eight sensors to the patient’s body with
the adhesive sleeves. The adhesive sleeves are supplied with
the Sensor Application kit.
To secure the Sensor Array cable to the patient’s hip, use the
adhesive pads. The adhesive pads are supplied with the
Sensor Application kit.
Note:
Any object which comes in between the patient’s body and
the sensors, including hair or air, and any changes in the
sensors arrangement may interfere with the quality of the
data.
!
To attach the Sensor Array
1.
Insert each sensor into an adhesive sleeve. The sensor
markings (dots or “THIS SIDE UP”) should face away
from the patient and the adhesive side of the sleeve
should face the patient’s abdomen.