User's Manual
Troubleshooting
Given Imaging Ltd.
98
If the LED does not blink upon Battery Pack
connection, refer to Data Recorder Problem #1. If
problem persists, reschedule for another examination
at a different date after excretion of the M2A™
Capsule, and contact Given® Imaging Limited
technical support.
If connection is intact:
1.
Change Battery Pack.
2.
Check blinking sequence of the Data Recorder LED:
If the blinking sequence is correct, proceed with
the examination and discharge the patient.
If the LED blinks, turns on and then off, but does
not blink according to the receiving rate,
reschedule for another examination at a different
date after excretion of the M2A™ Capsule.
If the LED does not blink upon Battery Pack
connection, refer to Data Recorder Problem #1. If
problem persists, reschedule for another examination
at a different date after excretion of the M2A™
Capsule, and contact Given® Imaging Limited
technical support.