User's Manual
Troubleshooting
Given Imaging Ltd.
97
battery is depleted. Consider completing the
examination.
If the LED does not blink upon Battery Pack
connection according to the blinking sequence,
refer to Data Recorder Problem #1.
PROBLEM #5
The Data Recorder LED stops blinking during the
examination and less than 6 hours have passed since the
Capsule ingestion.
SOLUTION
•
Tell the patient to keep the Data Record belt on, and heck
the Battery Pack-Data Recorder connection.
If there is a problem with the connection:
1.
Reconnect.
2.
Verify the blinking sequence of the Data Recorder
LED:
If the LED blinks upon Battery Pack connection
and then blinks according to the rate of
recording, continue with the examination.
If the LED blinks upon Battery Pack connection
but does not blink when receiving data, the
M2A™ Capsule is defective.
Reschedule for another examination at a different
date after excretion of the M2A™ Capsule.