User's Manual
Troubleshooting
Given Imaging Ltd.
96
depletion. (However, the belt must be worn in any event for a
minimum of 8 hours.). To check the connection and the
Battery Pack, perform the following:
•
Battery Pack and Data Recorder Connection
If there is a problem with the connection,
1.
Reconnect the Battery Pack to the Recorder
2.
Verify the blinking sequence of the Data
Recorder LED:
If the LED blinks upon Battery Pack connection
and then blinks according to the rate of
recording, continue with the examination.
If the LED does not blink upon connection of the
Battery Pack,, refer to Data Recorder Problem #1.
If the LED blinks upon Battery Pack connection
but does not start blinking according to the rate
of recording with an active capsule, the capsule’s
battery is depleted. Consider completing the
examination.
If the connection is intact,
1.
Change Battery Pack.
2.
Verify the Data Recorder LED blinking
sequence:
If the LED blinks according to the sequence
proceed with the examination.
If the LED blinks upon Battery Pack connection
but does not start blinking according to the rate
of recording with an active capsule, the capsule’s