User's Manual
Troubleshooting
Given Imaging Ltd.
95
•
Call technical support to report the faulty Data
recorder
PROBLEM #3
Data Recorder LED does not blink when the M2A™ Capsule
is in the patient’s mouth.
SOLUTION
1.
Tell the patient to keep the Data Record belt on, and
instruct the patient to take the M2A™ Capsule out of
his/her mouth.
2.
Hold the M2A™ Capsule next to a sensor attached to the
patient’s abdomen.
If the LED blinks, ask the patient to ingest the Capsule.
If not, replace the M2A™ Capsule and restart the
M2A™ Capsule Dispensing procedure as described on
page
84
. Be sure to dispose of the first capsule as per
the instructions in this manual.
PROBLEM #4
The Data Recorder LED stops blinking during the
examination and more than 6 hours have passed since the
Capsule ingestion.
SOLUTION
A minimum of 6 hours is considered a successful
examination. However, it is recommended to try and continue
with the examination, and to check that the termination is not
due to a connection fault or to a Battery Pack premature