User's Manual

Troubleshooting
Given Imaging Ltd.
94
PROBLEM # 2
The Data Recorder LED does not blink while the M2A
Capsule is held near the Sensor Array.
SOLUTION
1.
Move the M2A™ Capsule away from the Sensor Array.
2.
Disconnect the Battery Pack from the Data Recorder and
then reconnect them.
3.
Verify that the Data Recorder LED blinks and then turns
off.
If it blinks and then turns off, bring the M2A
Capsule close to the Sensor Array.
If the Data Recorder LED blinks, you may
proceed with the procedure of dispensing the
M2A Capsule.
If the Data Recorder LED does not blink, use
another M2A Capsule. Be sure to dispose of the
first capsule as per the instructions in this
manual.
If the Data Recorder LED does not blink, change the
Battery Pack.
If the Data Recorder LED blinks upon Battery
Pack connection, you may proceed with the
procedure of dispensing the M2A Capsule.
If the problem persists:
Change the Data Recorder and repeat the
Recorder Initialization process.