User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
PillCam Capsule Endoscopy
12 Accuracy of the Device—ESO
Accuracy of the Device—ESO
The PillCam Capsule Endoscopy System with the PillCam ESO 1 capsule was studied in a series of 107
subjects included in the study. Among patients included in this clinical study, 94 subjects had suspected
GERD at the time of the study, and a small group (13 subjects) were diagnosed with Barrett’s prior to the
study.
All patients underwent esophageal capsule endoscopy as well as esophagoscopy, to compare the results of
the esophageal capsule endoscopy with conventional video EGD (imperfect standard) in assessing
endoscopic findings of the esophagus.
1
The study was too small to accurately assess the agreement with lesions other than the most common
findings—esophagitis and suspected Barrett’s esophagus. There were several other lesions which were
noted as depicted in the table below.
Rate of
Agreement
(95% CI)
Rate of False
Pos (95% CI)
Rate of False
Neg (95% CI)
Overall
Agreement
(95% CI)
Esophagitis
85%
(69%; 95%)
13%
(4%; 27%)
12%
(3%; 27%)
Suspected BE
81%
(62%; 94%)
15%
(6%; 30%)
15%
(4%; 34%)
Normal
70%
(53%; 83%)
OVERALL
78.3%
(69%; 86%)
1.Clinical report presented in K041149
# Cases Diagnosed
by EGD
# Cases correctly
Diagnosed by CE
Agreement
Hiatal Hernia
37 6 16%
Stricture
5 1 20%
Polyps/Nodules
3 0 0%
Varices
1 1 100%
֠
֠֠
֠
Note
The PillCam ESO 1 capsule was not able to reliably identify:
• Precise location of lesion(s)
• Length or extent of lesion(s)
• Severity or grade of lesion(s)