User's Manual

Table Of Contents
Configuring RAPID Software
Regimen Manager 183
2. Click the General, Video, Recorder, Capsule & Sensor tabs and enter the information in the
editable fields. If there is a video that needs to be sent as well, click the Browse button in Video
logs and navigate to the video.
3. Click the Email button to send the data to Given Imaging Customer Support.
Regimen Manager
The Regimen Manager is a RAPID software tool for generating pre-capsule ingestion regimen
instructions to give to the patient before the day of the procedure and post-capsule ingestion regimen
instructions for the patient to follow during the capsule endoscopy procedure. The Regimen Manager
allows the creation and maintenance of a library of different instruction sets appropriate for different
circumstances to choose from. The post-capsule ingestion regimen instruction set is selected during
Check-in from the regimen library. The instruction set is uploaded to the PillCam recorder DR3 and is
used for creating timely alerts to the patient during the procedure. Currently only procedures using the
PillCam recorder DR3 with PillCam COLON 2 capsule utilize regimen instructions during the
procedure.
This section gives an overview of the Regimen Manager tool as well as the following:
Using the Regimen Manager on page 183
Pre-Ingestion Patient Instructions on page 185
Post-Ingestion Patient Instructions on page 187
Printing the Patient Instructions on page 190
Print Layout of the Post-Capsule Ingestion Instructions on page 190
Using the Regimen Manager
To access the Regimen Manager from the Home screen select Tools > Regimen Manager. The
Regimen Manager comes pre-loaded with several pre- and post-capsule ingestion regimens for you to
use. Any regimen to be used must be in an approved status by the user. None of the pre-loaded
regimens in the library are in an approved status. Any regimen in the library to be valid and selectable
in the menu must be reviewed and made approved by a physician. You can approve a regimen only
when it is opened and from within the opened regimen screen.