User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
Indications, Contraindications, Warnings, Cautions
Adverse Events 5
• In patients with dysphagia or other swallowing disorders.
• In patients with allergies or known contraindication to the medications and preparation agents used
in the procedure as described in the relevant instructions for use.
Adverse Events
Potential adverse events associated with the use of this device may include delayed or no excretion of
the capsule, aspiration, obstruction, perforation, and mucosal injury or bleeding. In some instances,
intervention is required to remove the capsule.
Warnings
Procedure Related:
• The absence of blood when performing an evaluation with the PillCam UGI video capsule does not
exclude the presence of a significant bleeding site in the acute upper gastrointestinal (esophagus,
stomach and duodenum).
• The PillCam COLON 2 capsule may be used for individuals after an incomplete optical
colonoscopy with adequate preparation, and a complete evaluation of the colon, was not
technically possible.
• A normal or negative capsule endoscopy examination does not exclude the possibility of colon
polyps or colon cancer.
• A negative or normal result obtained by the PillCam video capsule does not exclude the presence
of pathology and if symptoms persist, further evaluation should be performed.
• The safety of the PillCam SB capsule has not been established in children below two years of age.
• The safety of the PillCam ESO/PillCam UGI capsule has not been established in patients below
age 18.
• The safety of the PillCam COLON capsule has not been established in patients below age 18.
• In a small number of cases, the PillCam COLON capsule used for Colon capsule endoscopy may
not image the entire large bowel (colon) due to variations in patient GI motility. In the MA-204
study 104 (11.8%) subjects were excluded due to issues with the capsule procedure including 77
(8.7%) subjects with inadequate cleansing or short transit time for the capsule.
• If an adequate cleansing level is not achieved and the total transit time of the capsule is less than 40
minutes related, a repeat evaluation should be considered with the PillCam COLON capsule, or
with alternative diagnostic modalities.
• The capsule should not be swallowed by children under the age of 8 years or patients where a
concern for aspiration of the capsule exists (e.g., due to cognitive or neurological deficits or a
history of aspiration). In these patients, it is recommended that a capsule endoscopic delivery
!
Warning
PillCam capsules are MR unsafe.