User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
Performing PillCam Capsule Endoscopy
Capsule Ingestion 83
PillCam Recorder DR3
1. Make sure that the capsule and the capsule LED on the PillCam recorder are blinking in time
with each other.
2. Check the color of the capsule LED on the PillCam recorder:
• If (blue), go to step 3.
• If (white), pair the capsule with the PillCam recorder and verify the pairing success icon
appears in the top right corner of the PillCam recorder screen and the capsule LED blinks
blue .
3. Have the patient swallow the capsule with a sip of water. The ingestion procedure may take several
minutes.
PillCam Recorder DR2
1. Make sure that the capsule and the capsule LED on the PillCam recorder are blinking in time with
each other.
2. Position the patient on the bed, with a pillow (6 cm or 2.5 inch high) under the head to facilitate
drinking and ingestion.
3. Instruct the patient not to talk during the procedure.
4. To verify that the system is operating properly, hold the capsule in front of the patient's abdomen
very close to the sensors (practically touching through the clothes one of the sensors). Hold it for at
least 15 seconds and check that the capsule LED on the PillCam recorder blinks at the same rate as
the capsule.
5. Have the patient swallow the capsule with a sip of water.
֠
֠֠
֠
Note
During capsule endoscopy, the capsule LED blinks blue at the recording
rate.
If the capsule LED blinks orange once every 5 seconds, the PillCam
recorder is not receiving capsule signals (see PillCam Recorder DR3 on
page 209 for troubleshooting details).