Chapter 6
Performing PillCam Capsule Endoscopy
Setting Delay First Instruction in PillCam Recorder
Before ingestion of the PillCam COLON 2 capsule, you need to make sure that the settings match the
procedure circumstance: patient stays in clinic after ingestion or is instructed to go home immediately
after ingestion. If the patient stays in clinic, the
Delay first instruction flag on the PillCam recorder
DR3 needs to be OFF, otherwise it needs to be ON. If this flag has not been set properly during check-
in, it needs to be set before ingestion directly on the PillCam recorder prior to pairing with the PillCam
COLON 2 capsule by using the right navigation button with the icon above it to toggle the flag.
The default setting for a PillCam recorder DR3 checked-in with RAPID v8.3, without changing the
Delay first instruction field during check-in, is ON—first instruction delayed. Repetitive use of the
right button will toggle the status of the PillCam recorder DR3 and its status will show in the fourth
icon (from above) in the left most side of the fourth procedure data line.
The fourth line status icon and the button designation icon in the lower right corner of the screen above
the right navigation button will always be opposite. After pairing, the Delay first instruction icon will
disappear and the status will not be changeable any more.
Delay first instructionOff
Delay first instructionOn