User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
PillCam Capsule Endoscopy
76 PillCam Recorder DR2
These are the LED indicators and their statuses and colors for the most common PillCam recorder DR2
events and statuses:
LEDs PillCam Recorder DR2 Status
PillCam recorder is ON but not initialized. PillCam recorder does not capture
capsule signals.
PillCam recorder is initialized with patient data and ready to capture capsule
signals.
PillCam recorder shuts down if no capsule signals are received for more than
30, 60, or 90 minutes, depending on the PillCam recorder software version.
PillCam recorder is exchanging status or data with RAPID or RAPID RT.
LED blinking rate varies according to the communication flow.
PillCam recorder is capturing capsule signals.
Blinking rate = capsule frame rate.
PillCam recorder has stopped capturing capsule signals for more than 5
seconds.
PillCam recorder is detecting a capsule in sleep mode.
Blinking rate = every five seconds (in any color).
PillCam recorder is malfunctioning.
<20 seconds
PillCam recorder is synchronizing with a capsule. This is normal functioning.
>20 seconds
PillCam recorder detects capsule signal, but is not recording it. This is a
malfunction.
Check the sensor array connection or have patient move to a different location.
maximum Battery charge level:
When charging, the battery LEDs do not blink.
When PillCam recorder is out of the cradle, the battery
LEDS blink once every 5 seconds.
25%