76 PillCam Recorder DR2
These are the LED indicators and their statuses and colors for the most common PillCam recorder DR2
events and statuses:
LEDs PillCam Recorder DR2 Status
PillCam recorder is ON but not initialized. PillCam recorder does not capture
capsule signals.
PillCam recorder is initialized with patient data and ready to capture capsule
signals.
PillCam recorder shuts down if no capsule signals are received for more than
30, 60, or 90 minutes, depending on the PillCam recorder software version.
PillCam recorder is exchanging status or data with RAPID or RAPID RT.
LED blinking rate varies according to the communication flow.
PillCam recorder is capturing capsule signals.
Blinking rate = capsule frame rate.
PillCam recorder has stopped capturing capsule signals for more than 5
seconds.
PillCam recorder is detecting a capsule in sleep mode.
Blinking rate = every five seconds (in any color).
PillCam recorder is malfunctioning.
<20 seconds
PillCam recorder is synchronizing with a capsule. This is normal functioning.
>20 seconds
PillCam recorder detects capsule signal, but is not recording it. This is a
malfunction.
Check the sensor array connection or have patient move to a different location.
maximum Battery charge level:
When charging, the battery LEDs do not blink.
When PillCam recorder is out of the cradle, the battery
LEDS blink once every 5 seconds.
25%
below 10%