User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
Troubleshooting
Sensor Belt 209
Sensor Belt
Capsule
Cradle
PillCam Recorder DR3
Problem Cause Action
Connector is damaged
• Mishandling the sensor
belt
• End of Life
• Stop using this sensor
• Contact Customer Support
Insulation of the sensor wire
is damaged
Problem Cause Action
DOA (Dead On Arrival) Capsule failure 1 Send capsule to Given
Imaging Ltd.
2 Open another capsule.
3 If second capsule from 10-
pack is DOA, contact
Customer Support.
Problem Cause Action
Cradle LEDs turn red All LEDs are flashing red 1 Disconnect cradle from
mains power.
2 Reconnect cradle to mains
power.
3 If problem persists, contact
Customer Support.
Cradle orange LED is
blinking, but cradle is not in
discharge mode
Hardware/Software problem
PillCam recorder DR2 cannot
be placed in cradle
Hardware malfunction Contact Customer Support
Problem Cause Action
Capsule LED blinking red Interference
Recorder malfunction
Refer to PillCam recorder DR3 LEDs
Indications
Capsule LED blinking white Capsule not paired yet
SD card errors Wrong or damaged SD
card
Sensor connection Wrong or damaged
sensor
PillCam recorder freezes Possible static
electricity discharge
• Perform Hard Reset on the
PillCam recorder
• Contact Customer Support