User's Manual

Table Of Contents
RAPID Video 207
Appendix A4
Troubleshooting
RAPID Video
Saving and Opening Videos
Problem Cause Action
Short video Capsule
PillCam recorder battery
PillCam recorder
mishandling
Contact Customer Support
Send video on CD/DVD
Inform capsule lot#
Do not use the same PillCam
recorder
Save the raw data locally on
your computer
Gaps Capsule
Interference
Mishandling
Physiological
Bad image quality Stripes in video
Pixilation/confetti
Dark/red/orange image
Video shorter than
capsule operating time
without either ingestion
phase images or body
exit images
Capsule
PillCam recorder battery
Interference
Send video on CD/DVD
Save the raw data locally on
your computer
Contact Customer Support
No Localization Malfunction of the sensor
array
Wrong sensor chosen in
the check-in process
Do not use the sensor
Contact Customer Support
Cannot open RAPID
Atlas
Atlas installation incomplete
or incorrect
Reinstall Atlas
Save the raw data locally on
your computer
Contact Customer Support
Cannot view COLON 2
video using RAPID v7.0
COLON 2 video compiled in
RAPID v8.0 cannot be viewed
using RAPID v7.0
Download and use RAPID Reader
v8.0 to view video
Unable to perform
various operations
RAPID computer is “stuck”
typically after a prolonged
period in the ON state without
restarting
It is recommended to restart your
computer at least once a month to
ensure optimal performance
Problem Cause Action
Cannot locate video Video was not saved in
E:\Videos
Video was not created
Incorrect patient’s name
Search for video in RAPID
Work Directories or in Study
Manager (video creation)
Contact Customer Support