User Manual
12
Standard Weight Calibration
Depress [REZERO] and [←][T ] [T ] key ,entry to calibration mode, this time display windows display
calibration information. (SPAN SW must be ON)
INDICATOR: ● LIGHT ○ FLASH
1: ZERO 2: STA 3: NET 4: SET/HOLD 5: Kg 6: BT 7:AC
OPERATION DISPLAY INDICATOR REMARK
WEIGHT 1 2 3 4 5 6 7
Enter [←][ T ][ T ]while
depressing [REZERO].
888888 ● ● ●
CAL ● ● ● ● Display calibration information and
span DATA
15,000 ● ● ● ●
PRESS [←] and [↑] 5,000 ● ● ● ● Modication Sample Weight,
Depress [ ENTER ] to enter zero
calibration.
PRESS [ ENTER ] 5,000 ● ● ● ●
PRESS [ ENTER ] CAL 0 ● ● Depress [ ENTERR ] to enter zero
calibration.
------ ● ● Zero calibration
PRESS [ ENTER ] CAL SP ● ● ● Depress [ ENTER ] to enter span
calibration.
------ ● ● Span calibration.
5,000 ● ● Calibration is end then return to
weigh mode
MANINTENANCE
The product does not require any special maintenance. The technical life of the scale depends on its
use. Do not use water or liquid products to clean scale, juts use a soft cloth.
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specic recycling point for electric and
electronic equipment.